Full search:

escsis-man

Sir Wiki: A wonderful wonky wiki on Stable Isotope Ratio mass spectroscopy

Earth Systems Center for Stable Isotopic Studies

A YIBS center at Yale University

the ESCSIS Online Manual--or how to run what we got

please visit the instruments, prep, or people pages too


Please note: You need to be trained on the mass spectrometer and peripheral before running them. You should also examine the mass spec’s and peripheral’s user guides, appropriate application reports from Thermo Finnigan, P. A. de Groot’s “Handbook of Stable Isotope Analytical Techniques” that we have, the “Isogeochem” list server, and, of course, the literature for your area of study. This is only a guide, and is still a work in progress. If you got something to work well, please write down what you did so you can remember it--‘cause I’ll forget it.

1. File locations: Your data and sequence files are now located under:
Finnigan\stuff\<data, sequence, etc.>\<P.I.name>\<your-name>\
With your methods, results workshop, data export templates, etc. under the PI name only: Finnigan\stuff\<method, workshop, etc.>\<P.I.name>\
2. File paths in Isodat are set to data, method, etc., for appropriate heading and the upper most level will show the P.I. names. GC methods are a special issue, see section for GCC.
3. Template folders and files for methods and sequences are present. Please don’t change the template settings, but save them as “new name” (put in appropriate name) before making any changes.
4. Sequence set up:
5. Log in to log book for your instrument and put a print out of your sequence in the folder (three hold punch can be found in the mass spec room).
6. Load your samples, set the gases, check gas pressures on the tanks, let the system purge as needed, power up mass spec (select appropriate gas), check backgrounds as needed, and set system to remote control as needed (e.g. Costech EA).
7. If all ok and set up, push start and ok.
8. Wait for the first run to start to make sure all is ok and also check it at end of first run (in case Costech EA is not in remote mode or the GC is not set to auto-prep run, etc.)
9. Please NOTE: If the instrument and/or peripheral has not been used in a while, it is a very good idea to come down a day or two before you want to start your run with some junk samples so we can run some conditioning runs and make sure everything is behaving well.
10. At the end of a run, please take your print outs, e-mail or transfer the files from here to yourself (data and/or export files) so you have a copy of the data incase something happens down here, and can reprocess data if need be. Please do this before someone else starts their runs.

1. Sample Preparation:
2. Computer:
3. Mass Spec:
4. Check Instrument and Gas Configuration in Isodat:
High Voltage= 2.949 |Magnet=896
HV= 3.04* |
Box ~0.5 |Trap ~0.5 (may vary with current settings)
Vac ~1.4 x 10-6 mBar
Mass listed for N2:
28
29
30
The mV readings for the different masses may be higher at this time, depending on Conflo and Costech EA set up.
  • Please note, the HV should not be 0.3--click on “lightning” icon to turn HV back on if it’s at 0.3
5. Check Peripheral--Costech EA, Conflo III, and Gas tanks
6. Set-up Run
7. Start Run
8. Monitor first sample, then check back when a your first real sample is running. Follow to first run to make sure it is running OK (this is when you’ll realize you forgot to put the EA into remote mode). Check back later to make sure standards ran OK and then watch your first sample or two to make sure they’re fine too--this is especially important for new types of samples. If something is wrong, e.g. standards are off or it turns out the sample sizes are too large (or small), click on the red “stop” button to abort the sequence. It will ask you if you want to finish current sample--answer “yes” if the sample has dropped already.


9. End of Run--shut down:

1. Sample Preparation:
Typically we use 100-200ug silicate, 300-400 ug KF or 1-1.5mg PTFE, and run at a reduced flow rate. Remember, the KF is hydroscopic so work in a dry box if possible. Also, the PTFE + KF combo seem to work better than either alone. The sequence starts with blank (empty capsule) and reference (benzoic acid) runs, followed by a couple of conditioning/clean up runs (KF or PTFE alone), a fluorinated blank (KF or PTFE alone), then the samples. Please note, each sample usually requires one or two clean up runs (KF or PTFE alone) afterwards, and even then you may still see a memory effect. Fluoridated graphite (powder--don’t breath it in) can be used instead of PTFE, but it is much easier to slice off a fixed size of teflon tubing we have and place that into the silver capsule. Both PTFE and the fluoridated graphite are permeable to O2 from the air, which adds to the background. The KF must be dry. These runs should be done in a dedicated reaction tube, and a freshly packed tube will require extra conditioning/clean up runs. The crucible will have to be replaced after each PTFE sequence (49 runs) and every other KF sequence--tube cleaning with every second crucible change. Attempts to run at temperatures <1450 using PTFE did not do well. Initial tests combining KF with PTFE in a single run (say 200-400 ug each), seem to indicate they’ll work well together, as long as the KF is dry. Hydroscopic additives, KF, Ag F2?, etc., are difficult to work with without a dry box.
2. Computer:
3. Mass Spec, XP:
4. Check Instrument and Gas Configuration in Isodat:
High Voltage= 2.949 |Magnet=896
HV= 3.04* |
Box ~0.5 |Trap ~0.5 (may vary with current settings)
Vac ~1.4 x 10-6 mBar
Mass listed for N2:
28
29
30
The mV readings for the different masses may be higher at this time, depending on Conflo and Costech EA set up.
  • Please note, the HV should not be 0.3--click on “lightning” icon to turn HV back on if it’s at 0.3. The mV readings for the different masses may be high at this time, depending on Conflo and TC/EA set up. To double check to see if the conflo is opened up to the mass spec, pulse in Ref 1, CO, for a few seconds and see if the mass 28, 29 and 30 signals respond appropriately. Please contact the staff if you have to switch the mass spec to the conflo.
5. Check Peripheral--Thermo TC/EA, Conflo III, and Gas tanks
Once the H3+ factor has been calculated, lower Ref 2? again, wait a few seconds until the signal comes up, then set it to the desired intensity, like what was done with Ref 1? above. Please note, when lowering the pressure to lower the signal, the gas will take some time to bleed out. It it usually better to go below the desired value then raise the pressure up to bring the value up to the desired reading.
6. Set-up Run
7. Start Run
8. Monitor first sample, then check back when a your first real sample is running. Follow to first run to make sure it is running OK (this is when you’ll realize you forgot to put the EA into remote mode). Check back later to make sure standards ran OK and then watch your first sample or two to make sure they’re fine too--this is especially important for new types of samples. If something is wrong, e.g. standards are off or it turns out the sample sizes are too large (or small), click on the red “stop” button to abort the sequence. It will ask you if you want to finish current sample--answer “yes” if the sample has dropped already.

9. End of Run--shut down:

1. Sample Preparation:
Once the tubes are filled with 0.3% CO2 in He, the water is injected into the labelled tube. Inject 200 to 500 uL sample per tube: flush clean syringe with sample, load sample past mark and get rid of air bubbles, push plunger to mark and touch off any drops on needle tip, inject needle into to tube at one side, inject sample, touch off needle on side of tube, remove needle. Tube will be under slight positive pressure--that’s ok. If any droplets get on top of septa, wipe them off. Place tubes in order into heating block (temp 26o C). Wait 18-24 h. Please note waiting time is for pure water. Salty or dirty samples may take longer (days) to reach equilibrium. . ALSO: Condensate may have to be shaken down (sorry, no centrifuge yet) so it’s not on the septa. Do not let any water into the sampling syringe.
2. Computer:
3. Mass Spec, XP:
4. Check Instrument and Gas Configuration in Isodat:
High Voltage= 2.949 |Magnet=896
HV= 3.04* |
Box ~0.5 |Trap ~0.5 (may vary with current settings)
Vac ~1.4 x 10-6 mBar
Mass listed for N2:
28
29
30
The mV readings for the different masses may be high if the mass spec was recently opened to the Gasbench--or if there is a clog in the lines
  • Please note, the HV should not be 0.3--click on “lightning” icon to turn HV back on if it’s at 0.3.

5. Check Peripheral--Thermo Gasbench III and Gas tanks
6. Set-up Run
7. Start Run
8. Monitor first sample, especially line up of sampling syringe, then check back when a your first real sample is running. Follow to first run to make sure it is running OK. Check back later to make sure standards ran OK and then watch your first sample or two to make sure they’re fine too--this is especially important for new types of samples. If something is wrong, e.g. standards are off or it turns out the sample sizes are too large (or small), click on the red “stop” button to abort the sequence. It will ask you if you want to finish current sample--answer “yes” if the sample has dropped already.

9. End of Run--shut down:

1. Sample Preparation:
2. Computer:
3. Mass Spec:
4. Check Instrument and Gas Configuration in Isodat:
High Voltage= 9.005 |Magnet=5512
HV= 9* |
Box ~0.2 |Trap ~0.8(may vary with current settings)
Vac ~3.4 x 10-6 mBar
Mass listed for CO2: (expect reading <2-3mV)
44
45
46
The mV readings for the different masses may be high if the mass spec was recently opened to the GCC--or if there is a clog in the lines
  • Please note, the HV should not be 0.3--click on “lightning” icon to turn HV back on if it’s at 0.3.
5. Check Peripheral--Trace GC, GCC and Gas Tanks.


Helpful Hints from ....

aka, keeping things running....
Conflo:
1. Dilution doesn’t dilute/triggers apparent air leak.
GCC:
1.

Created on CREATEDATE 3/19/05 11:49 AM FILENAME ESCSIS-man.doc
modified for online wiki: 4/22/06
last updated: 4/22/06

Earth Systems Center for Stable Isotopic Studies, a Yale Institute for Biospheric Studies center
Dr. Danny Rye, ESCSIS Director; Dr. Gerard Olack, ESCSIS Manager

Welcome to WxWikiServer 1.5.11c!!



Links to this page: